| Term 
 | Definition 
 
        | any non-invasive technique to clinically evaluate the IOP by applying a foce to the globe and relating that force to teh deformation of the globe |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | the only method by which to get a true IOP measurement, however this is not a practical way to measure IOP since it involves the introduction of a cannula into teh eye |  | 
        |  | 
        
        | Term 
 
        | What are the three general types of tonometry? |  | Definition 
 
        | 1) Indentation (Schiotz) 2) Applanation (Goldmann, Perkins ahnd help, tonopen, pneumotonometer, mackay marg, NCT and Keeler pulsair) 3) Dynamic Contour Tonometry (PASCAL by Zeimer) |  | 
        |  | 
        
        | Term 
 
        | Describe indentation tonometry: |  | Definition 
 
        | Schiotz (rarely used anymore): 
measures the amount of indentation created by a fixed forceas the patient is lying down, the plunger of a set weight will indent the cornea by an amountthe amount of movement of the plunger indicaes the IOP by a reference chartthe method is similar to pushing your finger into a balloonsince this decreases the internal volume of the eye, it may give artificially high IOP |  | 
        |  | 
        
        | Term 
 
        | Describe applanation tonometry: |  | Definition 
 
        | Measures the force necessary to flatten a standard area 
Goldmann - pressure needed to applanate a standard areaPerkens hand held - standard areaTonopen - standard areaPneumotonometer - standard areaMackay Marg - standard areaNCT and Keeler Pulsair - amount of applanation with a standard force |  | 
        |  | 
        
        | Term 
 
        | Describe dynamic contour tonometry (DCT): |  | Definition 
 
        | provides a direct trans-corneal IOP measurement 
PASCAL by Zeimer:
claims to eliminate the errors with corneal thickness and rigiditycan also detect the ocular pulse amplitude (OPA) due to the patient's heartbeatmounted on the slit lampsingle use tip cover is needed |  | 
        |  | 
        
        | Term 
 
        | What is Imbert-Fick law?? |  | Definition 
 
        | the background theory for Goldmann tonometry: the pressure inside any sphere can be found by: W = P x A W = applanation force P = measured pressure A = applanation area   Considering this formula, the amount of pressure may be found by stnadardizing either the area of applanation (fixed area) or the aount of force (fixed force) |  | 
        |  | 
        
        | Term 
 
        | How must Imbert-Fick law be modified for the human cornea?  Why? |  | Definition 
 
        | the law only works if the membrane of the sphere is: 
infinitely thinperfectly flexibleperfectly elasticdry thus, the equation is modified for the cornea to: W + S = (P x Ai) +b S = force due to tears which pull the applanating surface towards the cornea b = force requred to bend the cornea due to its rigidity and thickness Ai - area of the inner corneal surface which is flattenered, since this is the surface against which the IOP act   NOTE: S anb b cancel each other out when teh applanation is performed with a 3.06mm diameter circle.  the Goldmann probes applanate a 3.06mm circle, therefore the simpler Imbert-Fick law is applicable to Goldmann tonometry |  | 
        |  | 
        
        | Term 
 
        | When should you perform tonometry? |  | Definition 
 
        | 
all patients during a routine eye exam for baseline dataapplanation tonometry should be performed on all cooperative patients regardless of ageall patients before dilation/cycloplegiaglaucoma suspects/patientsusually done at the end of the slit lamp exam and BEFORE dilation.  Hoever, man practitioners have NCT done as a pretest |  | 
        |  | 
        
        | Term 
 
        | When should you consider postponing tonometry? |  | Definition 
 
        | The following conditions relate to possibilities of infection or inaccuracy in measurement and may contraindicate tonometry: 
recent ocular injury (perforation, corneal abrasion or chemical burn)ocular infection or dischargeherpes on teh cornea or lidsmarked corneal edema: increass the chance of epithelial displacement and causes inaccurate readingsmarked nystagmuscorneal distortion, thickening, stromal loss or major scarringsignificant apprehension, blepharospasm, or strong Bell's phenomenon as avoidance responseconstant uncontrollable coughingendemic out breaking of ocular infections such as epidemic keratoconjunctivitis (adenovirus type 8)use your common sense for certain situations   |  | 
        |  | 
        
        | Term 
 
        | Inherent difficulties in using IOP as a test for glaucoma: |  | Definition 
 
        | 
ocular hypertensionlow tension and normotensive glaucomaupper limits of normal depends on the instrumentwide diurnal variationsinstrument variations unless IOPs are very high (40mmHg), glaucoma cannot be diagnosed based on IOP's alone |  | 
        |  | 
        
        | Term 
 
        | Expected Values: Normal
   
Normal:Most:Dirunal variations:Difference between OD and OS: |  | Definition 
 
        | 
Normal: 8-21 mmHgMost: 10-22 mmHgDirunal variations: 4-6 mmHgDifference between OD and OS: 2-3 mmHg |  | 
        |  | 
        
        | Term 
 
        | Expected  Values: Borderline
   
Borderline:Dirunal  variations:Difference  between OD and OS: |  | Definition 
 
        | 
Borderline: 21-24 mmHgDirunal   variations: 6-9 mmHgDifference   between OD and OS: 4-6 mmHg |  | 
        |  | 
        
        | Term 
 
        | Expected   Values: Abnormal
   
Abnormal:Dirunal   variations:Difference   between OD and OS: |  | Definition 
 
        | 
Abnormal: >24 mmHgDirunal   variations: >9 mmHgDifference   between OD and OS: >6 mmHg |  | 
        |  | 
        
        | Term 
 
        | IOP Measurement Variables: |  | Definition 
 
        | 
method of measurementdiurnal variationscorneal thicknessrefractive surgerypatient apprehension and tight lidsaccommodationfluid intakepharmaceutical agentscorneal hydrationcorneal biomechanics |  | 
        |  | 
        
        | Term 
 
        | Basics for ALL tonometry: |  | Definition 
 
        | 
tonometer must be clean, including all contact surfacesinstructions to patients and a brief description of what you will be doing |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
patient's namedoctor's namedateDPA's used and time instilledtime of tonometrytype of tonometryeach eye's recordings in mmHg (not an average) |  | 
        |  | 
        
        | Term 
 
        | Diagnostic Pharmaceutical Agents (DPA's):   What are they used for? |  | Definition 
 
        | 
used to aid in teh examination or assist in making a diagnosisnot used directly for treatmentanesthetics for tonometry   |  | 
        |  | 
        
        | Term 
 
        | What three things MUST be covered before installing ANY drops? |  | Definition 
 
        | 
Case History: Ocular History!
previous history with eye drops (any previous problems with drops? anesthetics?)pregnancy category CVA's (ALWAYS!!!)history of reacting to anesthetic at the dentist
consider alternative procedures which avoid use of anestheticsthis is very rare! |  | 
        |  | 
        
        | Term 
 
        | What color cap are local anesthetics? |  | Definition 
 | 
        |  | 
        
        | Term 
 
        | Uses for local anesthetics? |  | Definition 
 
        | 
tonometrygonioscopyexamining corneal injury if patient is unable to cooperate due to pain |  | 
        |  | 
        
        | Term 
 
        | Local Anesthetics:   
Onset:Duration:Effects on Cornea: |  | Definition 
 
        | 
Onset: 15 secondsDuration: 15-30 minutesEffects on Cornea: 
epithelium will remain "soft" for about one hourcornea is more permeable for up to 2 hours (make it more penetrable to teh mydriatic drops) |  | 
        |  | 
        
        | Term 
 
        | Local Anesthetics:   Side Effects |  | Definition 
 
        | 
delayed wound healingcorneal sloughing if used repeatedlydecrease fluorescence of dyessting upon applicationrednessfrontal headache |  | 
        |  | 
        
        | Term 
 
        | Local Anesthetics:   What are the two most commonly used? Which is the DOC?  Why? |  | Definition 
 
        | Proparacaine 0.5% 
DOCleast discomfortrare allergies   Tetracaine 0.5% 
more corneal toxicity than proparacainestings more than proparacainallergic reaction more common than with proparacaine |  | 
        |  | 
        
        | Term 
 
        | Describe non-contact tonometer: |  | Definition 
 
        | AO NCT: 
standardized air blast is directed at the corneacorneal deformity is measured by photoelectric cellsthis is transformed into a reading in mmHg |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
no anesthesiarelatively easy (?)less patient apprehension (??)little side effectseasy to train technicians to use |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
largeless accuratesome patients hate ita dinosaur |  | 
        |  | 
        
        | Term 
 | Definition 
 | 
        |  | 
        
        | Term 
 
        | Keeler Pulsair:   Advantages |  | Definition 
 
        | 
gentler puff (about 1/4 of AO NCT)great for kids! |  | 
        |  | 
        
        | Term 
 
        | Keeler Pulsair:   Disadvantages |  | Definition 
 
        | 
expensive (~$6,000-7,000)steep learning curve (tricky alignment until experienced)need about 5 readings to get a good averagemakes a noise like a vacuum cleaner which can be distracting to patients |  | 
        |  | 
        
        | Term 
 
        | Keeler Pulsair:   Instrument Parts |  | Definition 
 
        | 
the size of a large briefcasemain "case" has all teh pressure generating machinerycord to connect the housing and the hand held portion of the instrumentthe hand held portion looks like a staple gun |  | 
        |  | 
        
        | Term 
 
        | Keeler Pulsair:   Procedure |  | Definition 
 
        | 
educate the patient about the procedureswitch the machine onremove the hand piece from the cradle and press the set-reset switch to zero the machinehold the jet nozzle about 20mm from the patient's corneahold the tonometer plane to the patient's face and directly along the visual axishave the patient look between the two red lights and open widesteady the tonometer by placing your hand on both the instrument and the patient's foreheadalign the red corneal reflexes with the center of the eyepiecethe tonometer will activate automatically once the alignment is achievedpress the set-reset button and repeat the measurementrecord as with any other tonometry, stating the type of tonometry used |  | 
        |  | 
        
        | Term 
 
        | Hand Held NTC: (other than Keeler Pulsair)   Advantages |  | Definition 
 
        | 
portabilitycordlesseasy to use"softest puff available today"measures automatically upon alignment |  | 
        |  | 
        
        | Term 
 
        | Hand Held NTC: (other than Keeler Pulsair)   Procedure |  | Definition 
 
        | 
doctor views patient's eye through the eye piece and moves the instrument towards the patient's eyeonce the instrument has a view of an eye, an alignment system is activated automatically which guides the operatorit fires automatically once alignment is reached |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
smallgood for screenings and ambulatory patientsgood for patient's with corneal scarring |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
expensive (~$3,500)requires anestheticrequires tip covers |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
1.0mm transducer tipdisplays the average of 10 independent readingsdisplays statistical confidence of average readings |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
place latex cover over the probe prior to calibrationhold the switch down until a beep is heard, instrument will show = = =release teh switch and the display will be _ _ _ followed by another beephold the instrument vertically with the probe down and press the switch twice quickly.  Two beeps will sound and there will be a "CAL" displayanother beep will sound and the display with read "UP"now turn the tonometer so that the probe is up and the instrument is vertical.  A beep should sound followed by a readout of "Good".  You are now ready to startif "Bad" is displayed, repeate the above until "Good" is displayed |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
calibrate the instrument dailygive adequate patient instructionsinstill anestheticplace latex glove on the tonometer headfixation target for the patienthold instrument like a pen, perpendicular to the patient's central corneaanchor ring finger on patient's cheeklightly tap on the cornea several times listening for the chirp sounda beep will sound when enough valid readings have been takenan average IOP estimate will be displayedrecord your findings, indicating teh type of tonometry used |  | 
        |  | 
        
        | Term 
 
        | Proview Phosphene Tonometer:   Describe how it works |  | Definition 
 
        | 
uses pressure phosphene phenomenon to determine intraocular pressurepressure required to produce a phosphene correlates with IOP |  | 
        |  | 
        
        | Term 
 
        | Proview Phosphene Tonometer:   Advantages |  | Definition 
 
        | 
inexpensivewell toleratednon-invasiveself monitor/travel capabilities |  | 
        |  | 
        
        | Term 
 
        | Proview Phosphene Tonometer:   Disadvantages |  | Definition 
 
        | 
not well normedover estimates at higher IOPsfalse sense of securitynot everyone notices the phosphenes |  | 
        |  | 
        
        | Term 
 
        | Proview Phosphene Tonometer:   Procedure |  | Definition 
 
        | 
patient looks down and temporally (eye open)a small plunger is placed on the nasal lid and pressed until the patient notes the phosphenepressure is determined by how hard the plunger had to be presed |  | 
        |  | 
        
        | Term 
 
        | Icare Tonometer:   Advantages |  | Definition 
 
        | 
no anesthesia or dye are necessaryhand heldeasy to useportablewords for difficult patients |  | 
        |  | 
        
        | Term 
 
        | Icare Tonometer:   Disadvantages |  | Definition 
 
        | 
need a new probe for every patientexpensive (~$3,500) |  | 
        |  | 
        
        | Term 
 
        | Icare Tonometer:   Describe how it works |  | Definition 
 
        | 
uses a new measuring principle (Rebound Technology)a very light blunt tipped probe is projected onto teh corneathe probe rebounds off the cornea and sensors measure the IOP (average of 6 readings)the measurement is barely noticed by the patient (may not cause corneal reflex) |  | 
        |  | 
        
        | Term 
 
        | Diaton Tonometer:   Advantages |  | Definition 
 
        | 
measures IOP through the eyelidno anesthesia requiredno contact with teh corneahand held and portableBUT, expensive (~$2,500) |  | 
        |  | 
        
        | Term 
 
        | How is PASCAL DCT different than Goldmann? |  | Definition 
 
        | DCT is NOT a variable force tonometer |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   What method does it use? |  | Definition 
 
        | contour matching instead of applanation   (minimizes influencing factors, such as CCT, curvature, rigidity and elasticity) |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   Describe how the instrument works |  | Definition 
 
        | 
instrument uses a miniature pressure sensor embedded in the tipthe tip matches corneal curvature with minimal corneal deformation and constant oppositional forcemeasures IOP directly using the build-in pressure sensorywhen the sensor is subjected to a chance in pressure:
the electrical resistance is altered, andthe PASCAL's computer calculates a change in pressure similar to the change in resistance**this is the KEY to the PASCAL's ability to neutralize the effect of intra-individual variation in corneal propertiesonce the portion of the central cornea has taken up the shape of the tip, the integrated pressure sensor begins to acquire datameasures IOP 100 times per second (a complete measurement cycle requires about 8 seconds of contact time(during the measurement cycle, adio feedback is generated, which helps the clinician insure proper contact with the cornea |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   What is the key that allows PASCAL to neutralize the effect of intra-individual variation in corneal properties? |  | Definition 
 
        | when the sensor is subjected to a change in pressure: 
the electrical resistance is altered, andthe PASCAL's computer calculates a change in pressure similar to teh change in resistance |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   What is the design of the tip? |  | Definition 
 
        | 
tip radius: 10.5mmthe tip rests on the cornea with a constant appositional force of one gram
this is an important difference from all forms of applanation tonometry in which the probe force is variable |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   Details of measurements: |  | Definition 
 
        | 
allows for continuous reading of IOP waveformrecords the dynamic pressure fluctuations of the eyewhen the piezoresistive pressure sensor detects pressure, it is digitized and stored in teh device's memory
determines the intro-ocular pressure and its pulsatile fluctuations
OPA = ocular pulse amplitudecaused by cardiac activity |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
place new probe cover on sensor tippatient needs to be anesthetizedactivate it with a small turn of the blue knobadvance the probe to touch the eye 
need to have enough pressure to get adequate readingswill see a circular tear film area - center thisacquire proper soundneed to leave the tip on the cornea for at least 8 secondswhen the PASCAL sensor tip touches the corneal surface, a piezoresistive pressure sensor directly measures IOP |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   Digital Display |  | Definition 
 
        | 
IOP: average minimumOPA: ocular pulse amplitudeQ-Values: quality score of the reading (1-5)
1-2 = excellent3 = acceptable4-5 = unreliable |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   Potential Use of OPA |  | Definition 
 
        | 
normal OPA range found to be 3.0 +/- 1.2positive correlation between higher IOP and higher OPAlink between lower OPA and NTG (normotensive glaucoma)softer eye gives smaller OPA |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   OPA Correlations in Hispanic Population: |  | Definition 
 
        | 
OPA decreases with:
increasing axial lengthincreasing myopiaOPA increases with:
In this study, OPA:
range from 0.7-4.7 mmHgmean OPA was 2.1 mmHgFour subjects had OPA < 1.0 - all were myopic |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   Price of Instrument |  | Definition 
 
        | ~$6,000 range, which discount as you buy more of them   $0.75 per tip cover |  | 
        |  | 
        
        | Term 
 
        | Issues with GAT:   Corneal Thickness |  | Definition 
 
        | 
OHTS study found:
thicker corneas have higher measured IOP (false high) that doesn't respond well to medsthinner corneas have lower measured IOP (false low) and have a greater chance of converting from OHTN to glaucomaGAT measures based on 520 micrometers, but:
Caucasians: ave 555African-Americans: ave 535Mean K thickness: 545-550Need to find corrective factor for K thickness
studies have found anywhere from 1.5-4.5 mmHg for every 50 microns of corneal thicknesserror factor: 2.5 mmHg for every 50 microns away from 550 (subtract if above, add if below)LASIK patients:
1D of correction = 12 microns reduction in K5D correction = 60 microns reductin in Kdepending on corrective factor, need to adjust IOP anywhere from 1.5-4.5 mmHgadding pachymetry to the equation became the standard of care in the past few yearsrecent research is showing that this is an unreliable way to compensate for this problem |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   Reasons for PASCAL |  | Definition 
 
        | 
problems with GAT revealed by the OHTS study (adding pachymetry to equation of IOP is proving to be an unreliable way to compensate for varying corneal thicknesses)developed to accurately measure the IOP without the influence of corneal thickness and rigiditythe name is after Blaise Pascal, the seventeeth century mathematician and physicist |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
show that DCT IOP to closely compare to manometric values in cadaver eyessho that pre- and post-LASIK IOP reveals no significant change in IOP with the instrument |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   Corneal properties that affect IOP measurement: |  | Definition 
 
        | 
biomechanical material properties:
collagen and supporting matrixhydrationage (stiffer corneas and reduced tear film with age)effects of corneal biomechanical properties alone:
stiffer cornea = higher IOPsofter cornea = lower IOPpotential error is huge (>10 mmHg)corneal thickness (CCT)
thicker cornea = higher measured IOPthinner cornea = lower measured IOPpotential error is moderatecorneal curvature:
flatter cornea = lower measured IOPsteeper cornea = higher measured IOPpotential error is low |  | 
        |  | 
        
        | Term 
 
        | PASCAL DCT:   Summary of corneal biomechanics effects from Liu and Roberts: |  | Definition 
 
        | 
corneal radius of curvature: very smallcentral corneal thickeness: small/moderatecorneal material properties: large |  | 
        |  | 
        
        | Term 
 | Definition 
 
        | 
DCT is significantly less susceptible to the effects of CCT than GATGAT overestimates DCT in older subjects
suggests an age-related corneal biomechanical change that may induce errorwhere does that leave us? MORE RESEARCH! |  | 
        |  |